GSO ISO 14117:2024 Scope

This document specifies test methodologies for the evaluation of the electromagnetic compatibility (EMC) of active implantable cardiovascular devices that provide one or more therapies for bradycardia, tachycardia and cardiac resynchronization in conjunction with transvenous lead systems.

NOTE This document was designed for pulse generators used with endocardial leads or epicardial leads. At the time of this edition, the authors recognized the emergence of technologies that do not use endocardial leads or epicardial leads for which adaptations of this part will be required. Such adaptations are left to the discretion of manufacturers incorporating these technologies.

It specifies performance limits of these devices, which are subject to interactions with EM emitters operating across the EM spectrum in the two following ranges:

— 0 Hz ≤ ? < 385 MHz;

— 385 MHz ≤ ? ≤ 3 000 MHz

This document also specifies requirements for the protection of these devices from EM fields encountered in a therapeutic environment and defines their required accompanying documentation, providing manufacturers of EM emitters with information about their expected level of immunity.

Best Sellers From Health Sector

GSO 2017:2023
 
Gulf Standard
Personal Care and Baby Wet Wipes
GSO 2241:2023
 
Gulf Standard
Cosmetic Products - Body Cleaners
GSO 2063:2023
 
Gulf Standard
Cosmetic Products - Hair Conditioner (Balsam)
GSO 1223:2023
 
Gulf Standard
Cosmetic Products-Deodorants

Recently Published from Health Sector

GSO EN 14372:2024
EN 14372:2004 
Gulf Standard
Child use and care articles - Cutlery and feeding utensils - Safety requirements and tests
GSO ISO 10993-10:2024
ISO 10993-10:2021 
Gulf Standard
Biological evaluation of medical devices — Part 10: Tests for skin sensitization
GSO ISO 21649:2024
ISO 21649:2023 
Gulf Standard
Needle-free injection systems for medical use — Requirements and test methods
GSO ISO 15854:2024
ISO 15854:2023 
Gulf Standard
Dentistry — Casting and baseplate waxes