GSO ASTM F3306:2022 Scope
1.1 This test method assesses metal or other ions released
from single-use, metallic, implantable medical devices, or
components thereof, by exposing the device to solutions
simulating the in-vivo environment and temperature in a
container for a predetermined time frame with regular sampling at adequate intervals. Examples of device types that may
be evaluated by this test method include, but are not limited to:
cardiovascular devices, endovascular devices, and orthopedic
implants. Devices which are to be partially implanted, but in
long-term contact within the body (such as external fixation
devices) may also be evaluated using this method.
1.2 This test method is used to assess devices or device
components, or both, in their final form and finish, as they
would be implanted. For modular implants, consideration
should be given to individual testing of every part.
1.3 This test method covers the selection of specimens,
specimen preparation, test environment, method of exposure,
and method for evaluating the results to characterize ion
release. Because of the variety of configurations and sizes of
implants, a variety of specimen container configurations may
be used.
1.4 This test method is not intended for ions only adsorbed
onto the surface of the samples.
1.5 This test method does not apply to absorbable metallic
implants (for example, magnesium-based stents, bone screws,
etc.) that are intentionally designed to degrade in-vivo.
1.6 This test method does not cover the required subsequent
chemical analysis, for example, by inductively coupled plasma
mass spectrometry (ICP-MS), or the validation of the analytical instrumentation.
1.7 This test method does not cover the influence of
dynamic loading and associated surface damage or wear on ion
release. Ion release may change under dynamic loading conditions and wear. Additional testing may be required, depending on the application and outcome of this test method.
Best Sellers From Health Sector
GSO 2017:2023
Gulf Standard
Personal Care and Baby Wet Wipes
GSO 2241:2023
Gulf Standard
Cosmetic Products - Body Cleaners
GSO 1223:2023
Gulf Standard
Cosmetic Products-Deodorants
GSO 2156:2023
Gulf Standard
Cosmetic Products – Natural Henna
Recently Published from Health Sector
GSO ASTM F1148:2026
ASTM F1148:24
Gulf Standard
Standard Consumer Safety Performance Specification for Home Playground Equipment
GSO EN 71-18:2026
EN 71-18:2024
Gulf Standard
Safety of toys - Part 18: Phenol in aqueous (content) and polymeric (migration) toy materials
GSO 755:2026
Gulf Standard
Sterile cotton eye pad for medical purposes - Requirements
GSO ASTM F404:2026
ASTM F404:21
Gulf Standard
Standard Consumer Safety Specification for High Chairs