Activity Details
Radiological protection reference fields for calibrating dosemeters based on ISO 4037:2019
calibration measurements as per IAEA Safety Report Series No. 16:
1. Conditions and methods for producing reference X radiation:
· X ray Tube Potential Test
· Determine the inherent filtration for X-ray N-series
2. Field Uniformity and scattered radiation:
· Determination of the field size
· Determination of the field uniformity
· Determination of the scattered radiation
3. Validation of reference X radiation
· Criteria for validation by HVL determination
· Validation N-series qualities by HVL measurements
4. Establishment of reference gamma radiation
· Facility requirements, gamma irradiators, calibration bench, ancillary and dosimetry equipment
5. Calibration of Radiation Protection Dosimeter
· Fundamentals of calibration and verification
· Reference conditions and standard test conditions
· Methods of calibration based on IAEA safety report series No. 16
· The conversion coefficient for area dosimeter and personal dosimeter based on ISO 4037-3:2019
· Examples: Determine the air kerma rate and the calibration coefficient
· Examples: Determine the calibration coefficient for electronic personal dosimeter
· Records and certification